This application note highlights the electrochemical detection of copper ions (Cu²⁺) using differential pulse voltammetry (DPV), showcasing a reliable method for trace copper ion analysis in environmental and industrial samples with the Medpstat. The study assesses the stability, sensitivity, and reproducibility of the instrument for trace-level Cu²⁺ analysis in the micromolar concentration range. In addition, analytical performance is quantified through the determination of the limit of detection (LOD) and limit of quantitation (LOQ). The results demonstrate the robustness of the Medpstat and its suitability for reliable heavy-metal detection.
Introduction:
Monitoring heavy metals, particularly copper ions (Cu²⁺), remains a critical task in environmental electrochemistry, industrial process monitoring, and public health applications. Elevated copper concentrations in aqueous and soil matrices can lead to toxicity, whereas accurate quantification at trace levels is essential for regulatory and quality assurance purposes. Among the various analytical approaches, electrochemical techniques such as differential pulse voltammetry (DPV) are highly valued for their exceptional sensitivity, rapid response, and robust quantitation of metal ions at low concentrations.
In this application note, the electrochemical detection and quantification of Cu²⁺ ions are demonstrated at micromolar levels (5, 10, 30, 50, and 100 µM) using DPV measurements. The approach is further extended to determine the concentration of an unknown Cu²⁺ sample, providing a practical example of quantitative analysis. The limit of detection (LOD) and limit of quantitation (LOQ) are also evaluated, emphasizing the analytical performance, signal resolution, and quantitative reliability of the DPV-based electrochemical method for trace copper analysis.
What You Will Learn from This Application Note
- Why do you choose DPV techniques?
- What does the supporting electrolyte do?
- Why are LOD and LOQ important in ion detection?
Principle:
Differential pulse voltammetry (DPV) is an electroanalytical technique in which small voltage pulses are applied over a gradually increasing potential. The current is measured immediately before and after each pulse, and the difference (i₁ – i₂) is plotted against potential. By minimizing background (capacitive) currents, this approach produces sharp, well-defined peaks that are directly proportional to analyte concentration, offering high sensitivity and resolution for trace-level metal detection.
Reagents:
Working solutions of Cu²⁺ (5, 10, 30, 50, and 100 µM) were prepared in 0.1 M H₂SO₄ by diluting a 0.001 M CuSO₄ stock solution. For each case, the required stock volume (N1) was calculated using V1N1=V2N2, where V1=0.001 M, V2=5–100 µM, and N2= 10mL. The required stock volumes for 10 mL of each concentration are summarized in Table 1.

Why is the supporting electrolyte H2SO4, and not PBS?
An acidic supporting electrolyte (0.1 M H₂SO₄) was selected over PBS because acidic media provide higher conductivity and maintain Cu²⁺ in a free, stable ionic form. This results in sharper, more intense, and reproducible DPV peaks, whereas PBS at neutral pH offers lower conductivity, broader peak shapes, and can cause speciation issues that reduce sensitivity.
Apparatus

Electrochemical measurements were performed using the MedPstat portable potentiostat operated via MedEplot software (Windows compatible). A conventional three-electrode configuration was employed in a Mini Gas-Tight Cell 10ml (MTX Labs, India), comprising a Glassy Carbon (3mm) as working electrode, an Ag/AgCl Reference Electrode (6 mm diameter), and a Platinum Coil Counter Electrode (30 mm length). Electrodes were rinsed with deionized water and wiped before use. Connections to the potentiostat were made as follows: working electrode (red), reference electrode (blue), counter electrode (black), and sensing (red).
Parameters

Result and Discussion
A 200 µM CuSO₄ solution was freshly prepared by dissolving CuSO₄·5H₂O in 0.1 M H₂SO₄, which served as the supporting electrolyte. Before each measurement, the Glassy Carbon Electrode (GCE) was polished using a 0.05 µm alumina slurry on a brown microcloth pad and then rinsed thoroughly with deionized water, followed by gentle drying. This procedure ensured a clean, reproducible, and well-defined electrode surface for reliable measurements.

Cyclic voltammetry (CV) was performed within a potential window of +1.0 V to –0.6 V at a scan rate of 50 mV/s (Figure 2). The CV curve clearly revealed the electrochemical behavior of copper: a distinct anodic peak corresponding to copper oxidation was observed at approximately +0.1 V, and a cathodic peak associated with copper reduction appeared at –0.5 V. Beyond –0.6 V, water reduction was evident. These results confirm the successful detection of copper redox activity in the electrolyte.

Differential pulse voltammetry (DPV) measurements were conducted for copper ion concentrations in the range of 5–100 µM (Figure 3). The voltammograms displayed a distinct anodic peak at approximately 0.1 V, which can be attributed to the oxidation of copper. The peak current exhibited a linear dependence on concentration, confirming a clear concentration-dependent electrochemical response. As expected, the 100 µM solution produced the highest current signal, while progressively lower concentrations yielded proportionally smaller peak intensities. The DPV experiments were carried out under the following conditions: quiet time, 1 s; potential window, –0.6 to 0.5 V with a step increment of 5 mV. These optimized parameters provided stable baselines and highly reproducible responses, thereby enabling sensitive detection of Cu²⁺ (detailed parameter settings are presented in Figure 1).

To evaluate analytical performance, the peak currents obtained from DPV (Figure 2) were extracted and plotted against concentration (Figure 4). A linear regression fit was applied. The resulting calibration curve exhibited excellent linearity with a correlation coefficient (R²) of 0.99. and the slope along with the standard deviation of the response were used to calculate the limit of detection (LOD) and limit of quantification (LOQ).
The limit of detection (LOD) and limit of quantification (LOQ) were calculated using the following standard formulas:

Where,
σ: Standard deviation of the blank
S: Slope of the calibration curve
LOD (3.3): Detectable signal ~3× noise (~99% confidence)
LOQ (10): Quantifiable signal ~10× noise for reliable measurement

The calibration analysis revealed an LOD of approximately 1.0 ppm and an LOQ of approximately 4.8 ppm CuSO₄ in 0.1 M H₂SO₄. These results demonstrate that the developed electrochemical method is highly sensitive, allowing the detection of copper at trace levels (~1 ppm) even on a bare GCE without any surface modification. Reliable quantification is achieved at concentrations above 3 ppm, with solutions of 4.8 ppm (30 µM) and higher showing well-defined and proportional responses. Notably, even at lower concentrations of 0.8 ppm (5 µM) and 1.6 ppm (10 µM), distinct copper signals were observed, highlighting the method’s capability to detect trace copper efficiently.
Conclusion
The developed electrochemical method demonstrates high sensitivity for copper detection using a plain glassy carbon electrode without any special surface treatment, achieving a limit of detection (LOD) of approximately 1.6 ppm and a limit of quantification (LOQ) of approximately 4.8 ppm. The key to this performance is the MedPStat potentiostat, which enables measurements to be conducted easily, accurately, and reliably. Even at low concentrations, clear copper signals were observed, confirming the method’s capability for trace-level detection. Furthermore, combining the MedPStat with surface modifications or nanostructured electrodes could further enhance sensitivity, potentially reaching sub-ppm or nanomolar detection levels. This simple yet powerful approach highlights the method’s strong potential for environmental monitoring and trace metal analysis.